CCR傳感器您現(xiàn)在所在的位置: 首頁 > CCR傳感器
      • CCR 4H2S-100 工業(yè)硫化氫傳感器
      • CCR 4H2S-100 工業(yè)硫化氫傳感器
      • CCR 4H2S-100 工業(yè)硫化氫傳感器

      CCR 4H2S-100 工業(yè)硫化氫傳感器

      類 別: CCR傳感器

      描 述: ★ Fast response and recovery time ★ High stability ★ Robust at humidity and temperature extremes ★ Low cross sensitivity to methanol

      24小時服務熱線
      400-891-8898
      產(chǎn)品詳情


      Performance Characte ristics

      Output signal

      700 ± 250 nA / ppm

      Typical Baseline Range (pure air)

      ±2 ppm CO equivalent

      T90 Response Time

      < 30 seconds

      Measurement Range

      0 - 100 ppm

      Maximum Overload

      500 ppm

      Linearity

      Linear

      Repeatability

      < ±2% H2S equivalent

      Recommended Load Resistor

      10 ohms

      Resolution (Electronics dependent)

      < 0.1 ppm typical

      Environmental

      Temperature Range Continuous

      -30oC to +50oC

      Pressure Range

      800 to 1200 mbar

      Operating Humidity Range

      15% to 90% RH

      Lifetime

      Long Term Output Drift

      < 5% per annum

      Recommended Storage Temp

      0oC to 20oC

      Expected Operating Life

      > 24 months in air

      Standard Warranty

      24 months from date of dispatch

      Intrinsic Safety Da ta

      Maximum at 2000 ppm

      0.3 mA

      Maximum o/c Voltage

      1.3 V

      Maximum s/c Current

      <1.0 A


      Cross -Sensitivity Data

      Cross -Sensitivity Dat

      GAS

      CONC.

      GS+4H2S

      Carbon Monoxide

      100 ppm

      <2 ppm

      Sulphur dioxide

      20 ppm

      0 ppm

      Nitrogen Dioxide

      5 ppm

      <0.5 ppm

      Nitric Oxide

      50 ppm

      <0.5 ppm

      Ammonia

      50 ppm

      0 ppm

      Chlorine

      15 ppm

      0 ppm

      Ethylene

      100 ppm

      0 ppm

      Carbon Dioxide

      5000 ppm

      0 ppm

       

      Important Note: The values above are typical values and should not be used as a basis for cross calibration. Cross sensitivities may not be linear and should not be scaled either. Above data based on gassing for 5 minutes using CCR Scientific test equipment. Should be noted some cross interference break through will occur if gas is applied for a longer period of time.

      Poisoning:

           CCR Scientific sensors are designed to operate in a wide range of harsh environments and conditions. However, it is important that exposure to high concentrations of solvent vapors is avoided, both during storage, fitting into instrument and operation.

      相關產(chǎn)品
      立即訂購
      產(chǎn)品名稱:
      聯(lián)系人:
      電話:
      備注:
      驗證碼:
      相關新聞
    1. 上一篇:CCR R-4OX 工業(yè)氧氣傳感器
    2. 下一篇:沒有了;